Given Imaging Limited GIVENRECORDER Manual De Usuario

Descargar
Página de 80
PillCam Platform
56
Chapter 5
Voltage dips, 
short interruptions 
and voltage 
variations on 
power supply 
input lines
IEC 61000-4-11
<5 % U
T
 
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
 
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
 
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
 
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
 
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
 
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be 
that of a typical commercial or 
hospital environment. If the 
user of the DataRecorder 2 
requires continued operation 
during power mains 
interruptions, it is 
recommended that the 
DataRecorder 2 be powered 
from an uninterruptible power 
supply or a battery.
Power frequency 
(50/60 Hz) 
magnetic field, 
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical 
commercial or hospital 
environment.
NOTE: U
T
   is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration - electronic emissions
The DataRecorder 2 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the DataRecorder 2 should assure that it is used in such an environment.
Immunity test IEC 60601 test 
level
Compliance 
level
Electromagnetic environment - 
guidance
Portable and mobile RF communications 
equipment should be used no closer to 
any part of DataRecorder 2, including 
cables, than the recommended separation 
distance calculated from the equation 
applicable to the frequency of the 
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 VRMS
150 kHz to
80 MHz
3V
ms
d = 1.2 P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1.2 P   80 MHz to 800 MHz range
d = 2.3 P   800 MHz to 2.5 GHz range
Guidance and manufacturer’s declaration - electronic emissions