PillCam Platform
58
Chapter 5
DataRecorder 3 
Guidance and manufacturer’s declaration - electromagnetic emissions
The DataRecorder 3 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the DataRecorder 3 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Data Recorder 3 uses RF energy only for its 
internal function. Therefore, its RF emissions are very 
low and are not likely to cause any interference in 
nearby electronic equipment
RF emissions
CISPR 11
Class B
The Data Recorder 3 is suitable for use in all 
establishments, including domestic establishments 
and those directly connected to the public low-voltage 
power supply network that supplies buildings used for 
domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
Guidance and manufacturer’s declaration - electromagnetic immunity for all 
equipment and systems
The DataRecorder 3 is intended for use in the electromagnetic environment specified below. The customer 
or the user of the DataRecorder 3 should assure that it is used in such an environment.
Immunity test IEC 60601 test level
Compliance level
Electromagnetic 
environment - 
guidance
Electrostatic 
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the relative 
humidity should be at least 
30 %.
Electrical fast 
transient / burst
IEC 61000-4-4
2 kV for power supply lines
1 kV for input/output lines
N/A
Mains power quality should be 
that of a typical commercial or 
hospital environment.
Surge,
IEC 61000-4-5
1 kV line to line
2 kV line to earth
N/A
Mains power quality should be 
that of a typical commercial or 
hospital environment.