- 8 - 
 
Polarity 
The Polarity parameter allows the user to change the polarity of 
the output signal. 
 
 
 
 
 
Gain 
The Gain parameter allows the user to adjust the gain or scale 
factor applied to the input signal and is reflected in the display 
output, the Voltage Output signal, and the digital (RS-232) output 
signal.  The Transmitter Gain is displayed on the left ("Gain 
T:02000  S:02000") and is changed using the RM10K Remote 
Control.  The System Gain is shown on the right ("Gain T:02000 
S:02000") and is the parameter adjusted on the RX10K.   
 
 
 
 
 
The Transmitter Gain is set based on the microstrain (
µe) range 
expected during a given test (see table below; strain values 
assume a Gage Factor of 2.0).   
 
Transmitter Gain 
Setting 
Full Scale Strain 
(Full Bridge, 
µe) 
500 
±4000 
1000 
±2000 
2000 
±1000 
4000 
±500 
8000 
±250 
16000 
±125 
 
Full Scale Strain corresponds to a display reading of 10000, an 
analog Voltage Output value of 10 VDC, and a digital data output 
of 10000.  The System Gain can be adjusted from 25% to 400% of 
the Transmitter Gain (i.e., ¼ to 4 times the Transmitter Gain) for 
convenient output scaling.  The equation below relates strain input 
to Voltage Output. 
 
 
+00000 -_
 
_
 
_
 
_
 
_
 
 
_
 
_
 
_
 
_
 
_+ 
P
olarity:     
P
ositive 
+00000 -_
 
_
 
_
 
_
 
_
 
 
_
 
_
 
_
 
_
 
_+ 
G
ain 
T
:02000  
S
:02000 
TX
const
 
System 
Gain 
RX10K 
Voltage 
Output (V) 
Measured 
Strain (
µe) 
Gage 
Factor 
 x 
x