The Transmitter constant (TX
const
) is 2.5V. The Gage Factor is
specified by the gage manufacturer but is typically about 2.0. See
Appendix B for Full Scale torque and output scaling equations and
sample calculations.
Output Offset
The Output Offset allows the user to adjust the offset or “move the
zero” of the output from the RX10K. The adjustment value
displayed on the right is the actual output offset value in millivolts.
The adjustment range is from –12000mV to +12000mV (
Appendix B for Full Scale torque and output scaling equations and
sample calculations.
Output Offset
The Output Offset allows the user to adjust the offset or “move the
zero” of the output from the RX10K. The adjustment value
displayed on the right is the actual output offset value in millivolts.
The adjustment range is from –12000mV to +12000mV (
±12V),
meaning the zero can be moved anywhere within the output
range.
This adjustment affects the display output, the Voltage Output
signal, and the digital (RS-232) output signal. The Output Offset
value is applied to the signal after the Gain adjustment; therefore,
the Gain adjustment may affect the zero output signal.
User Default
The RX10K parameters can be returned to their default settings.
Holding down the ADJUST
range.
This adjustment affects the display output, the Voltage Output
signal, and the digital (RS-232) output signal. The Output Offset
value is applied to the signal after the Gain adjustment; therefore,
the Gain adjustment may affect the zero output signal.
User Default
The RX10K parameters can be returned to their default settings.
Holding down the ADJUST
▲ key while powering up the RX10K
resets the RX10K to the default values listed below.
Default
Default
Description
Rx Ch
1
Input
Transmitter
Filter
500Hz
Input AutoZero
Off
Polarity
Positive
Gain
T=S
Output Offset
0