The Transmitter constant (TX
const
) is 2.5V.  The Gage Factor is 
specified by the gage manufacturer but is typically about 2.0.  See 
Appendix B for Full Scale torque and output scaling equations and 
sample calculations. 
 
Output Offset 
The Output Offset allows the user to adjust the offset or “move the 
zero” of the output from the RX10K.  The adjustment value 
displayed on the right is the actual output offset value in millivolts.  
The adjustment range is from –12000mV to +12000mV (
±12V), 
meaning the zero can be moved anywhere within the output 
range. 
 
 
 
 
 
This adjustment affects the display output, the Voltage Output 
signal, and the digital (RS-232) output signal.  The Output Offset 
value is applied to the signal after the Gain adjustment; therefore, 
the Gain adjustment may affect the zero output signal.  
 
User Default 
The RX10K parameters can be returned to their default settings.  
Holding down the ADJUST 
▲ key while powering up the RX10K 
resets the RX10K to the default values listed below. 
 
Default 
 
Description 
Rx Ch   
 
Input 
 
 
Transmitter 
Filter 
 
 
500Hz 
Input AutoZero   
Off 
Polarity  
 
Positive 
Gain 
 
 
T=S 
Output Offset 
 
 
+00000 -_
 
_
 
_
 
_
 
_
 
 
_
 
_
 
_
 
_
 
_+ 
0
utput 
0
ffset: +00000