Boston Scientific Corporation CRM330017 Manual De Usuario

Descargar
Página de 80
51
4.
Unplug all equipment cables from the side panels of the LATITUDE
Programming System.
NOTE: See each accessory’s product literature for transport and storage
conditions. Ensure each accessory is maintained within the appropriate limits.
Long Term Storage of the LATITUDE Programming System
If the Programmer is to be stored for long periods (e.g. months), remove the
battery to prevent it from discharging to a point where recharging would be
required in order to use it again. See "Battery Status, Installation,
Replacement, and Recycling" on page 45 
for instructions on removing the
battery.
Maintenance Check and Safety Measures
LATITUDE Programming System Maintenance Check
Prior to each use, you should perform a visual inspection and verify the
following:
Mechanical and functional integrity of the LATITUDE Programming
System, cables, and accessories.
Legibility and adherence of the LATITUDE Programming System labels.
Perform "Startup" on page 24. The normal power-up process verifies that
the LATITUDE Programming System has passed its internal checks and is
ready for use.
NOTE:
The LATITUDE Programming System does not contain any user-
serviceable parts and it does not have a calibration requirement. Maintenance
does not require any additional steps.
The LATITUDE Programming System contains only one user-accessible
component, the Model 6753 replaceable Lithium-ion battery.
NOTE:
The Programmer must be returned without the battery for
replacement or repair of any internal components. See "Battery Recycling" on
page 49 
for additional details.
Safety Measurements
National regulations may require that the user, manufacturer, or manufacturer
representative periodically perform and document safety tests of the device. If
such testing is required in your country, follow the testing interval and extent of
testing as regulated in your country. If you do not know the national regulations
in your country, please contact your local Boston Scientific representative.
It is not necessary that technical and safety inspections be performed by
Boston Scientific personnel. However, technical and safety inspections of the
Programmer and its accessories must be performed by persons, who, based
on their training, knowledge, and practical experience, are capable of
adequately performing such inspections and who do not require instructions
with regard to the technical and safety inspection.
If IEC/EN 62353 is a required standard in your country, but no specific testing
or interval is specified, it is recommended that you perform safety testing using