Table 3.
Possible Causes and Corrective Actions for LATITUDE Programming
System Problems (continued)
Symptom
Possible Cause
Corrective Action
the back cover of this
manual to confirm PG and
Model 3300 Programmer
compatibility.
Incorrect telemetry wand
Use only the Model 6395
Telemetry Wand for
transvenous PGs.
Poor connection between
a telemetry wand and the
Programmer
Disconnect and reconnect
the Model 6395 Telemetry
Wand to the Programmer.
Programmer running on
battery power only
Connect the Programmer
to AC power to improve
telemetry performance.
Excessive radio emissions
from equipment
Reposition the LATITUDE
Programming System.
Also see Noise problems:
ECG.
Incomplete telemetry
communication with Model
6395 Telemetry Wand
Reposition the Model
6395 Telemetry Wand
over the PG; repeat
interrogation.
Flip the wand over.
Disconnect and reconnect
the wand. Turn off the
Programmer, and then
turn it On. Repeat
interrogation.
Use another Model 3300
Programmer or Model
6395 Telemetry Wand.
Repeat interrogation.
If this does not correct the
issue, contact Boston
Scientific using the
information on the back
cover of this manual.
Telemetry RF signal
obstructed
Assure that a clear line-of-
sight path exists between
the LATITUDE
Programming System and
PG. Repeat interrogation.