Q-TECH™ PROGRAMMER:  OPERATiOn
Using the Wand
Model 3203 wand (“the wand”) makes it possible for this programmer to communicate with the 
SQ-RX device.
Note:   The wand is a non-sterile device. DO NOT STERILIZE the wand or the 
programmer. The wand must be contained in a sterile barrier before use 
in the sterile field.
To connect the wand to the programmer, slide the wand cable connector over the communication 
connector port located on the rear edge of the programmer (Figure 5)
To disconnect the wand, grasp the wand cable connector and gently pull it straight off the 
communication connector port.
Note:   Do not pull or yank on the cable to disconnect the wand from the 
programmer. Such action could cause hidden damage to the cable. A 
damaged cable might reduce wireless communication capabilities and 
require a replacement wand. 
Figure 5: Connecting the wand
To obtain optimal telemetry, the wand should be placed directly over the implanted SQ-RX device. 
Although the programmer may receive S-ECG telemetry at greater distances, programming should 
be performed with the wand placed over the SQ-RX device.