Boston Scientific Corporation CRM70514 Manual De Usuario

Descargar
Página de 86
53
Q-TECH™ PROGRAMMER:  OPERATiOn
Capture and View S-ECG Strips
The programmer can display, capture and store real-time S-ECG rhythm strips. The programmer 
saves a maximum of fifteen, 12-second, Captured S-ECG recordings comprised of:
• 
8.5 seconds before activation of the Capture S-ECG button
• 
3.5 seconds after activation of the Capture S-ECG button
If an additional recording is required, then the oldest previous recording is replaced with the new 
recording.
To capture a new S-ECG rhythm strip:
1. 
Select the Capture S-ECG button located to the left of the Live S-ECG window 
(Figure 43). The S-ECG will scroll across the display screen. Calipers appear below the 
Captured S-ECG rhythm strip (Figure 44). Each 12-second recording is date and time 
stamped according to the programmer’s date and time setting.
Figure 43:  Select the Capture S-ECG button
Figure 44:  Captured S-ECG rhythm strip