Q-TECH™ PROGRAMMER: OPERATiOn
Capture and View S-ECG Strips
The programmer can display, capture and store real-time S-ECG rhythm strips. The programmer
saves a maximum of fifteen, 12-second, Captured S-ECG recordings comprised of:
•
•
8.5 seconds before activation of the Capture S-ECG button
•
3.5 seconds after activation of the Capture S-ECG button
If an additional recording is required, then the oldest previous recording is replaced with the new
recording.
To capture a new S-ECG rhythm strip:
1.
Select the Capture S-ECG button located to the left of the Live S-ECG window
(Figure 43). The S-ECG will scroll across the display screen. Calipers appear below the
Captured S-ECG rhythm strip (Figure 44). Each 12-second recording is date and time
stamped according to the programmer’s date and time setting.
Figure 43: Select the Capture S-ECG button
Figure 44: Captured S-ECG rhythm strip