2
Q-TECH™ PROGRAMMER: GEnERAl dEsCRiPTiOn
•
The US power cord is for connection to 110 VAC supply mains. All other power cords are for
connection to 230 VAC supply mains. Outside North America, use the supplied power cord that
exactly matches your AC electrical outlet.
•
To avoid risk of electric shock, the programmer must only be connected to a supply mains with
protective earth.
•
Mains isolation is achieved by disconnecting the AC power cord. Do not position the
programmer or the external power supply in a manner that would make it difficult to
disconnect that cord.
•
Never use a damaged external power supply or a damaged programmer.
•
The programmer is not waterproof or explosion-proof and it cannot be sterilized. Do not use
it in the presence of flammable gas mixtures, including anesthetic mixtures containing air,
oxygen, or nitrous oxide.
•
Confirm that the programmer is in communication with the intended implanted SQ-RX Pulse
Generator.
•
The programmer is intended for use by healthcare professionals only. The programmer
may cause radio interference or may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures, such as reorienting or relocating the programmer or
shielding the location.
•
By design, the programmer emits radio frequencies in the 402-405 MHz and 2.4 GHz bands.
This may interfere with nearby equipment. Closely monitor medical electrical equipment in the
vicinity when using the programmer.
•
The presence of other equipment in the same frequency bands (402-405 MHz for the
SQ-RX Pulse Generator and 2.4 GHz for the printer) may interfere with communication. If
communication problems occur, refer to the Troubleshooting section of this manual.
•
If the programmer is used near other medical or office equipment, all equipment should be
observed to verify normal operation. Other equipment may interfere with the programmer,
even if that other equipment complies with the International Special Committee on Radio
Interference (CISPR) emission requirements.
•
Use of non-Cameron Health/Boston Scientific-approved accessories may result in increased
emissions or decreased functionality of the programmer and may cause unintended
operational behavior of the programmer.
•
No modification of the programmer is allowed.
•
Use only the Cameron Health/Boston Scientific S-ICD programmer and appropriate software to
communicate with and program the SQ-RX Pulse Generator.
•
Do not discard the programmer in a fire, incinerate, or subject it to temperatures that exceed
100° C (212° F). This could cause the programmer to explode.