Q-TECH™ PROGRAMMER:  GEnERAl dEsCRiPTiOn
• 
Placing a magnet over the SQ-RX Pulse Generator suspends arrhythmia detection and therapy 
response. Removing the magnet resumes arrhythmia detection and therapy response.
• 
Battery depletion will eventually cause the SQ-RX Pulse Generator to stop functioning. 
Defibrillation and excessive numbers of charging cycles shorten the SQ-RX Pulse Generator’s 
battery longevity.
• 
The S-ICD System has not been evaluated for pediatric use.
• 
The S-ICD System does not provide long-term bradycardia pacing, Cardiac Resynchronization 
Therapy (CRT) or Anti-Tachycardia Pacing (ATP).
Use of Other Medical Therapies/Diagnostic Procedures
• 
External defibrillation or cardioversion may damage the S-ICD System. Avoid placing the 
defibrillation paddles directly over the SQ-RX device or electrode.
• 
Cardio Pulmonary Resuscitation (CPR) may temporarily interfere with sensing and may cause 
delay of therapy.
• 
Do not expose a patient with an implanted S-ICD System to diathermy. The interaction of 
diathermy therapy with an implanted SQ-RX Pulse Generator can damage the SQ-RX Pulse 
Generator and cause patient injury.
• 
Do not expose the patient to MRI scanning. MRI scanning can damage the SQ-RX Pulse 
Generator and cause patient injury.
• 
Electrical interference or “noise” from sources such as electrosurgical and monitoring 
equipment can interfere with the communication between the programmer and SQ-RX Pulse 
Generator or cause inappropriate therapy. If interference occurs, move and reposition the 
programmer or wand away from the source of the interference.
• 
Ionizing radiation therapy, such as radioactive cobalt, linear accelerators, and betatrons 
may adversely affect the S-ICD System operation. Therapeutic ionizing radiation cannot be 
immediately detected; however, it can damage the electronic components of the SQ-RX Pulse 
Generator. To minimize the risks of ionizing radiation:
 
»
Shield the SQ-RX Pulse Generator with a radiation-resistant material, regardless of the 
distance between the SQ-RX Pulse Generator and the radiation beam.
 
»
Do not project the radiation port directly at the SQ-RX Pulse Generator.
 
»
Evaluate the S-ICD System operation after each radiation treatment.
• 
Lithotripsy and other therapeutic forms of ultrasound can damage the SQ-RX Pulse Generator. 
If required, avoid direct flow of the pulse waves near the site of the implanted SQ-RX Pulse 
Generator.