Boston Scientific Corporation CRMU22814 Manual De Usuario

Descargar
Página de 40
RA
RV
IS-1
UNI/BI
IS-1
UNI/BI
[1] RA [2] RV [3] Suture Hole
Figure 5. Lead connections and setscrew locations, RA: IS-1, RV: IS-1
NOTE: The pulse generator case is used as a pace electrode when the pulse generator has been
programmed to a unipolar lead setting.
IMPLANTING THE PULSE GENERATOR
Implant the pulse generator by performing the following steps in the sequence provided. Some patients may
require pacing therapies immediately upon connecting the leads to the pulse generator. If modifications to the
nominal settings are needed, consider programming the pulse generator before or in parallel with implanting
the lead system and forming the implantation pocket.
Step A: Check Equipment
It is recommended that instrumentation for cardiac monitoring, defibrillation, and lead signal measurement
should be available during the implant procedure. This includes the PRM system with its related accessories
and the software application. Before beginning the implantation procedure, become completely familiar with the
operation of all the equipment and the information in the respective operator’s and user’s manuals. Verify the
operational status of all equipment that may be used during the procedure. In case of accidental damage or
contamination, the following should be available:
56