Sterile duplicates of all implantable items
Sterile wand
Sterile PSA cables
Torque and non-torque wrenches
During the implantation procedure, always have a standard transthoracic defibrillator with external pads
or paddles available for use.
Step B: Interrogate and Check the Pulse Generator
To maintain sterility, test the pulse generator as described below before opening the sterile blister tray. The
pulse generator should be at room temperature to ensure accurately measured parameters.
1.
Interrogate the pulse generator using the PRM. Verify that the pulse generator’s Device Mode is
programmed to Storage. If otherwise, contact Boston Scientific using the information on the back cover.
To begin a ZIP telemetry session for ACCOLADE, PROPONENT, and ESSENTIO devices, verify that the
ZOOM Wireless Transmitter is connected to the PRM via the USB cable and that the green light on top of
the transmitter is illuminated. To initiate communication with all devices, position the wand over the PG
and use the PRM to Interrogate the pulse generator. Keep the telemetry wand in position until either a
message appears, indicating that the telemetry wand may be removed from proximity of the pulse
generator, or the ZIP telemetry light illuminates on the PRM system. Select the End Session button to
quit a telemetry session and return to the startup screen. Radio frequency interference may temporarily
disrupt ZIP telemetry communication. Increasing the distance from the source of interfering signals or
repositioning the ZOOM Wireless Transmitter may improve ZIP telemetry performance. If ZIP telemetry
performance is not satisfactory, the option of using wanded telemetry is available.
2.
Review the pulse generator’s current battery status. Counters should be at zero. If the pulse generator
battery status is not at full capacity, do not implant the pulse generator. Contact Boston Scientific using
the information on the back cover.