CAUTION:
  Use only the Power Module supplied with the EPIC 10 system (Biolase 
 
                    Part Number 6400142).
 
Access Key Code 
The Access Key Code prevents unauthorized use of the system. It is activated every 
time system is turned on with the Power Switch.  
 
NOTE:
  Turning the laser off by pressing and holding the Control button on the front   
 
             panel does not re-set the Access Key Code.  Turn the Power Switch OFF only 
 
             when the system will not be in use for a long period of time.
 
 
CONTROL Button 
Once the power switch is set to the ON position the access key code is activated.  The 
CONTROL button on the control panel must be pressed to enable the footswitch.  The 
aiming beam will illuminate to indicate that the system is ready for use.  
 
Wireless Footswitch 
The 
EPIC 10
 will not emit laser energy until the user presses down on the footswitch in 
READY mode. The footswitch is designed to work using wireless technology.  Two AAA 
batteries are required to power the footswitch (operating voltage – 2 to 3.2 volts). 
 
   
                                                       Figure 3: Footswitch
 
 
Remote Interlock 
This feature allows the device to be connected to the remote sensor which will prevent 
its operation when triggered (e.g., by opening door). The electric cable from this 
connector should be wired to the normally closed switch, sensing the opening of a door 
and turning the laser OFF when switch is open.  
 
This feature is overridden when the plug is not connected. 
 
       
 
                                                                  
Figure 4: Remote Interlock connector