TrueCPR Device Instructions for Use 
Using the TrueCPR Device
recommended that routine post-event confirmation of implantable device 
function be performed by a trained clinician to ensure programming and system 
integrity.
Using the TrueCPR Device for Training
The TrueCPR device may be used on a manikin for training purposes. Some 
manikins may require updates to reach compression depths of 5 to 6 cm (2 to 
2.5 in), consistent with AHA and ERC guidelines. You can use a manikin that 
does not allow these depths, but the TrueCPR device will not reach the 
compression target zone.
Note:  
manikin contains large metal plates.
Screen Illustrations
Calibration Screen
When you turn on the TrueCPR device, the calibration screen appears. 
Compression Indicators
During compressions, the gray fan moves between the recoil zone and 
compression zone as shown.
Calibration in progress
Calibration complete
Compression in 
progress: Gray fan 
moves across screen.
Sufficient depth: Dark 
green wedge in 
compression target 
zone.
Sufficient recoil: Dark 
green wedge in recoil 
target zone.