TrueCPR Device Instructions for Use 
Data Management
Data Management
The TrueCPR device can store compression data for three 60-minute sessions, 
or up to six sessions totaling 180 minutes. When all available memory has been 
used, the data from the oldest use is overwritten automatically. 
Data can be transferred to a computer with TrueCPR device-compatible software 
via USB connection. Event reports can be printed directly from the software. For 
information on TrueCPR device reports and full event review functionality, 
contact your local Physio-Control representative. 
To initiate data transfer: 
1
Connect a USB cable between the TrueCPR device and a computer with 
compatible software.
Note: 
Use only a USB cable that meets the following specifications: USB 2.0 
A-Male to Mini-B 5-pin Male, 28/24 AWG cable with ferrite core (gold plated), 
1 m (3 ft). 
2
Turn on the TrueCPR device. If the software on the computer does not start 
automatically, open the program manually.
3
Follow the instructions in the software on the computer to select the TrueCPR 
device and initiate data transfer. Data transfer may take up to 5 minutes if 
memory is full. The following screens may appear on the TrueCPR device:
WARNING
SHOCK HAZARD: Do not use the USB connector if the TrueCPR device 
is damaged.
POSSIBLE ELECTRICAL INTERFERENCE: Using cables or accessories 
that are not specified for use with the TrueCPR device may result in increased 
emissions or decreased immunity from electromagnetic or radio frequency 
interference (RFI) for this device. This could affect the performance of the 
TrueCPR device or equipment in close proximity. Use only cables and 
accessories specified in these instructions for use.
Data transfer in 
progress.
Data transfer successful.
Data transfer failed.