Wave Gain Tab (Patient)
Patient Setup Functions
5 - 52
0070-10-0690-01
Panorama™ Operating Instructions
5.6
Wave Gain Tab (Patient)
The Wave Gain tab (shown in FIGURE 5-33) is used to change the wave gain settings for
the selected patient’s available waveforms. The layout, functionality, and buttons available in
this tab will change based on the type of monitoring device assigned to the selected patient
tile.
the selected patient’s available waveforms. The layout, functionality, and buttons available in
this tab will change based on the type of monitoring device assigned to the selected patient
tile.
NOTE:
Changing the ECG wave gain does not affect the arrhythmia
algorithm.
algorithm.
The Wave Gain tab is available at the ViewStation and the WorkStation, and functions in
the same manner as it does at the host central station. When a patient tile is assigned at the
ViewStation or the WorkStation, it will initially display the wave gain settings of the
associated host central station patient tile, provided the patient tile is communicating with the
central station.
the same manner as it does at the host central station. When a patient tile is assigned at the
ViewStation or the WorkStation, it will initially display the wave gain settings of the
associated host central station patient tile, provided the patient tile is communicating with the
central station.
After this initial transfer, changes made to the host central station, the ViewStation and/or the
WorkStation patient tile Wave Gain tab will be independent of each other unless
communication is interrupted. Automatic updating of the wave gain settings at the remote
workstation patient tile will occur if any of the following occur:
WorkStation patient tile Wave Gain tab will be independent of each other unless
communication is interrupted. Automatic updating of the wave gain settings at the remote
workstation patient tile will occur if any of the following occur:
• The monitoring device is placed in the Standby mode at the central station
• The monitoring device is not communicating with its host central station (Communications
Lost)
• Communication between the host central station and the ViewStation and/or the
WorkStation is lost (Central Communications Lost)
The default host central station wave gain settings are defined by the System Wave Gain
tab, which is described in the “Wave Gain Tab (System)” on page 9-59.
tab, which is described in the “Wave Gain Tab (System)” on page 9-59.
5.6.1
Accessing the Wave Gain Tab (Patient)
1. From the menu bar, select the Patient Setup menu button. The tabs associated with
patient setup are displayed.
2. Choose a patient tile by selecting its digital tile or waveform area.
3. Select the Wave Gain tab. The Wave Gain tab is enabled.
FIGURE 5-33 Wave Gain Tab