Patient Alarm Setup Tab (Optional)
Patient Alarm Functions
6 - 2
0070-10-0690-01
Panorama™ Operating Instructions
6.1
Patient Alarm Setup Tab (Optional)
The Patient Alarm Setup tab (shown in FIGURE 6-1) provides password protected access 
to the patient alarm configuration. When the correct password is entered, the Alarm Limits 
and the Alarm Responses tabs are opened. 
The Patient Alarm Setup tab is only shown if the Password Protection option is set to ON 
in the System Alarms tab. If Password Protection is OFF, the Patient Alarms menu 
button provides direct access to the Alarm Limits and the Alarm Responses tabs. For 
additional information, refer to the ‘‘System Alarms Tab’’ on page 9-14.
The Patient Alarm Setup tab is available at the central station and the WorkStation. 
6.1.1
Accessing the Patient Alarm Setup Tab 
From the menu bar, select the Patient Alarms menu button. The Patient Alarm Setup 
tab is displayed.
FIGURE 6-1 Patient Alarm Setup Tab 
6.1.2
Entering a Password
To enter a password:
1. Select the Password text box in the Patient Alarm Setup tab. The keyboard dialog 
box is displayed. 
2. Use the keyboard dialog box to enter the Alarm password that was created in the 
Passwords tab (maximum 15 characters). For security purposes asterisks (*) are 
shown for each typed character. For additional information regarding passwords, refer 
to the ‘‘Passwords Tab’’ on page 9-29 or contact your System Administrator.
NOTE:
Passwords are case sensitive. 
•  The Backspace (<==) button clears a single character in the edit box 
• The Caps Lock button enables/disables the caps lock function. A Caps Lock 
indicator is shown when Caps Lock is enabled. 
• The Close button discards the changes
• The Enter button accepts changes