Mindray DS USA Inc. WT0707-00000 Manual De Usuario

Descargar
Página de 484
Panorama™ Operating Instructions
0070-10-0690-01
6 - 7
Patient Alarm Functions
Alarm Limits Tab (Patient)
6.2.3
Simultaneous Alarm Limit Editing
If using a bedside device with the Panorama Central Station, alarm limit and alarm response 
configuration changes cannot be made at the bedside monitor, central station, and 
WorkStation simultaneously. The bedside device will take precedence, followed by the 
central station, ending with the WorkStation.When an attempt is made to make changes 
simultaneously, the message These settings are currently being edited remotely.will 
be displayed at the central station or WorkStation, depending on where the change was 
initiated. If there are multiple WorkStations on the network, the first WorkStation that initiates 
the change will take precedence over any other WorkStation attempting to implement the 
same change.
These are the scenarios in which this message will be displayed: 
• The Alarm Limits Menu is active at the bedside monitor and then the patient Alarm 
Limits tab is opened at the central station or the WorkStation
• The Alarm Limits tab is active at the central station or the WorkStation and then the 
Alarm Limits Menu is opened at the bedside monitor 
In both of these scenarios, unsaved changes made in the Alarm Limits tab will be 
discarded.
To close this message dialog, select the OK button. Acknowledging this message will convert 
the Alarm Limits tab at the central station to a read-only state and editing will not be 
permitted. Once the Alarm Limits Menu is closed at the bedside device, the Alarm Limits 
tab will become active again and editing can resume.
For additional information regarding the bedside monitor menus, refer to bedside monitor’s 
Operating Instructions for the Passport 2
®
 (P/N 0070-00-0649-XX), Spectrum
 (P/N 
0070-00-0648-XX), or Spectrum OR
 (P/N 0070-00-0670-XX).
6.2.4
Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Restore Previous Settings 
Select the Restore Previous Settings sidebar button to return patient display settings to 
those in use when the tab was opened.
NOTE:
The Restore Previous Settings sidebar button is not available 
at the ViewStation.
Delta Temp (F°)
25%
Delta Temp (C°)
25%
TABLE 6-1 Auto Set Percent Indicator (continued)
ALARM
PERCENT INDICATOR
(APPROXIMATE)