Alarm Limits Tab (Patient)
Patient Alarm Functions
6 - 8
0070-10-0690-01
Panorama™ Operating Instructions
Suspend All Alarms (Optional)
Select the Suspend All Alarms sidebar button to suspend, for a user configured period of 
time, all numeric physiological, lethal arrhythmia, non-lethal arrhythmia, and technical 
alarms at the Panorama Central Station for the selected patient tile. The Suspend All 
Alarms
 sidebar button is only shown if the Suspend Alarms option is ON in the System 
Alarms
 tab. 
NOTE:
The Suspend All Alarms sidebar button is not available at 
the ViewStation.
NOTE:
The amount of time an alarm is suspended is dependent on 
the configured System Suspend All Timeout setting.
When the Suspend All Alarms sidebar button is selected:
• The Suspend All Alarms sidebar button changes to white with black text.
• The All Alarms Suspended message displays in red text in the Patient Status line of 
the patient tile. For additional information regarding the Patient Status line, refer to the 
‘‘Patient Status Line’’ on page 2-10.
To resume the monitoring of alarms, select the Suspend All Alarms sidebar button.
• When the Suspend All Alarms sidebar button is disabled, the button changes to blue 
with white text. Alarms are now re-enabled for the selected patient.
Suspend Non-Lethal 
Select the Suspend Non-Lethal sidebar button to suspend all numeric, technical and 
non-lethal arrhythmia alarms for the selected patient tile. Lethal arrhythmia alarms continue to 
be monitored at the Panorama Central Station when non-lethal arrhythmia alarms are 
suspended.
NOTE:
The Suspend Non-lethal sidebar button is not available at 
the ViewStation.
NOTE:
The Brady alarm will not be suspended when the Suspend 
Non-Lethals sidebar button is selected.
When the Suspend Non-Lethal button is selected:
• The Suspend Non-Lethal sidebar button changes to white with black text.
• The Non-Lethal Alarms Suspended message displays in yellow text in the Patient 
Status line of the patient tile. For additional information regarding the Patient Status line, 
refer to the ‘‘Patient Status Line’’ on page 2-10.
To resume the monitoring of non-lethal alarms, select the Suspend Non-Lethal sidebar 
button.