Panorama™ Operating Instructions
0070-10-0690-01
8 - 5
Report Functions
Patient Reports Tab
• A parameter label
• A numeric measurement
• A unit of measure label 
• A heart rate (HR) value 
• Dashes (--) to indicate missing or invalid digital data. 
• Square brackets ( [ ] ) to indicate alarming parameter values
NOTE:
Digital Data is only shown on the first page of the report.
8.1.2.3
Waveform Data 
The waveforms that print in the All Strips Report’s waveform data section depend on which 
leads are currently in use and on the waveforms selected in the Print Setup function. 
The first page of this report contains four waveforms. The middle pages of the report will 
contain six waveforms while the last page may contain one to six waveform strips depending 
on the waveforms available. Each waveform in this report represents approximately 10 
seconds of data; five seconds prior to the time of the print request and 5 seconds after the 
print request. A waveform label (i.e., ECG Lead II) is shown above each strip to identify the 
data source and the unit of measurement (i.e.,%, RPM, BPM, mmHg, kPa). Pleth, ECG and 
Respiration do not display a unit of measurement. The waveforms displayed in the report are 
shown on a calibrated grid and contain a scale bar. 
If the All Strips Report printed as a result of a Print on Alarm, the event that prompted the 
report will print below the report digital data in bold. 
For pacemaker patients that have Pacer Enhancement enabled, the report displays a vertical 
bar to indicate the firing of the pacemaker impulse. 
For information on selecting the waveforms that can be printed in this report, refer to the 
"Print Setup Tab (Patient)" on page 5-55.
8.1.2.4
ECG Waveforms
The ECG waveforms shown in the All Strips Report are based on the ECG lead wire set 
currently in use and on the waveforms selected in the Print Setup tab. The ECG waveform 
report displays the following:
• An ECG lead label
• A wave gain scale bar and a calibration pulse for wave gains less than 30 mm/mV
• A  calibrated  grid
• Current ECG monitoring label
• A continuous straight line for active ECG leads that are not connected 
Non-ECG Waveforms
The non-ECG waveforms shown in the All Strips Report only include the non-ECG parameters 
in use at the time of the print request.