Panorama™ Operating Instructions
0070-10-0690-01
1 - 7
System Overview
Supported Devices
1.3.7
Passport 2
®
 Bedside Monitor
Passport 2 bedside monitors are used to monitor patient data and may use either a 
hardwired or a wireless (2.4 GHz or 608 MHz) configuration. The Passport 2
 
sends
 
data
 
to 
the Panorama Central Station for analysis and storage via an Ethernet connection or 
remotely via the Panorama Telemetry Server. The data collected by the Passport 2 is based 
on the type of lead wire set and sensors used. 
If an alarm in the following table does not have an associated footnote, it may occur using 
any lead wire set. 
Parameters/Waveforms
PARAMETERS
SETTINGS
PANORAMA ALARMS
ECG/Heart Rate
Leads I, II, III are available with a 3-wire 
lead set.
Low Heart Rate
High Heart Rate
Leads I, II, III, aVR, aVL, aVF, V are 
available with a 5-wire lead set.
Leads I, II, III, aVR, aVL, aVF, V1, V2, 
V3, V4, V5, V6 are available with a 
View 12
 card.
(View 12 supported by Passport 2 
hardwired and WMTS 608)
ST
ST analysis is in use at the bedside 
device. 
ST Single
ST Dual
PVC/min.
Arrhythmia analysis is in use at the 
bedside device.
Asystole 
Bigeminy
Brady
2
Couplet
Irregular Heart Rate
2
 
Pause
1
High PVC
Run
Trigeminy
V-Fib
V-Rhythm
V-Tach
Respiration (Source: 
ECG, CO
2
 and Gas
3
 
Module)
Respiration is in use at the bedside and 
is derived from a valid source. 
Low Respiration
High Respiration
Apnea
SpO
2
SpO
2
 sensor is in use at the bedside 
device.
Low SpO
2
 
High SpO
2
 
IBP1/IBP2 (Systolic, 
Diastolic, Mean)
IBPx option is in use at the bedside 
device.
IBP parameter labels include IBP, Art, 
UA, LV, PA, CVP, ICP, LA and RA. The 
label used is based on the label selected 
at the bedside monitor.
Low IBPx Systolic
High IBPx Systolic
Low IBPx Diastolic
High IBPx Diastolic
Low IBPx Mean
High IBPx Mean
NOTE:
x = 1 or 2
INSP/ET CO
2
Gas Module is in use at the bedside 
device. 
High CO
2
 Inspired
Low CO
2
 End Tidal
High CO
2
 End Tidal
1
Alarm will only be detected when using a View 12
 card.
2
Alarm will not be detected when using a View 12
 card.
3
Waveform not supported when using a WMTS 608 bedside device.