Mindray DS USA Inc. WT0707-00000 Manual De Usuario

Descargar
Página de 484
Unit Choices Tab
System Setup Functions
9 - 52
0070-10-0690-01
Panorama™ Operating Instructions
3. Select the Pacer Reject button to select the system default pacer reject period. Pacer 
Reject adjusts the time period at which pacemaker-induced ECG complexes are filtered 
for ECG sampling processes. 
Pacer Reject options include 0 to 80 milliseconds (ms) in 20-second increments. The 
factory default setting is 40 ms.
NOTE:
Changing the Pacer Reject option will affect all newly 
admitted patients and will have no affect on patients 
currently being monitored.
NOTE:
The Pacer Reject option is not available at the ViewStation 
or the WorkStation.
4. Select the Paging Demographics button to define which demographic information is 
sent to the paging system when a page is requested. Paging Demographics options 
include BedBed + Last Name, and Bed + ID. The factory default setting is Bed
Refer to the following tables for formatting examples:
NOTE:
The amount of data that is displayed in the pager display 
window may differ slightly due to the type of pager being 
used. 
NOTE:
Changing the Paging Demographics option will take effect 
immediately. 
OPTION
ALARM TYPE
FORMAT EXAMPLES 
Bed
Heart Rate 
central name:XXXBed:ICU10 Low Heart Rate Alarm 
Bed
Arrhythmia 
central name:XXXBed:ICU10 Asystole 
Bed
Not HR or 
arrhythmia alarm
central name:XXXBed:ICU10 High Temperature 1 
Bed
ST alarm
central name:XXXBed:ICU10 ST Single Alarm 
OPTION
ALARM TYPE
FORMAT EXAMPLES 
Bed + Last 
Name
Heart Rate 
central name:XXXBed:ICU10 Name:Patient Last Name 
Low Heart Rate Alarm
Bed + Last 
Name
Arrhythmia 
central name:XXXBed:ICU10 Name:Patient Last Name 
Asystole 
Bed + Last 
Name
Not HR or 
arrhythmia alarm
central name:XXXBed:ICU10 Name:Patient Last Name 
High Temperature 1 
Bed + Last 
Name
ST alarm
central name
:
XXXBed:ICU10 Name:Patient Last Name ST 
Single Alarm 
OPTION
ALARM TYPE
FORMAT EXAMPLES 
Bed + ID
Heart Rate 
central name:XXXBed:ICU10 ID:Patient ID Low Heart Rate 
Alarm 
Bed + ID
Arrhythmia 
central name:XXXBed:ICU10 ID:Patient ID Asystole 
Bed + ID
Not HR or 
arrhythmia alarm
central name:XXXBed:ICU10 ID:Patient ID High 
Temperature 1 
Bed + ID
ST alarm
central name
:
XXXBed:ICU10 ID:Patient ID ST Single 
Alarm