Mindray DS USA Inc. WT0707-00000 Manual De Usuario

Descargar
Página de 484
Telepack with Nonin SpO2 Module (Optional)
Telepack (Optional)
10 - 14
0070-10-0690-01
Panorama™ Operating Instructions
2. Route the cable in the channel on the back of the unit. The channel regions (shown in 
FIGURE 10-12) should also be pressed firmly to ensure that they are properly placed.
FIGURE 10-12 Securing the Patient Cable
10.5.5
Attaching the Finger Clip Sensor
1. Insert the finger (preferably the index, middle, or ring finger) into the Nonin finger clip 
until the end of the finger reaches the finger stop (shown in FIGURE 10-13). Keep the 
fingernail facing the sensor top. Ensure that long fingernails are not interfering with the 
proper finger position.
NOTE:
Disposable finger sensors (shown in FIGURE 10-14) may 
also be used. Follow the packaging directions carefully to 
ensure proper application.
NOTE:
For best results, when using the sensor for data collection, 
secure the sensor cable independently from the sensor with 
medical device tape, preferably around the base of the 
finger. Be sure that the tape securing the cable does not 
restrict the blood flow. 
CAUTION:
FIGURE 10-13 Reusable finger clip 
sensor
FIGURE 10-14 Disposable sensor