Panorama™ Operating Instructions
0070-10-0690-01
3 - 3
Alarm Behavior
Physiological Alarms
Once the alarm condition is resolved and/or acknowledged, the audible alarm stops, unless 
the alarm is latched. For additional information regarding alarm limits, refer to ‘‘Alarm Limit 
Settings’’ on page 13-2
. For additional information regarding alarm latching, refer to 
‘‘Latching’’ on page 3-3.
3.2.2
Alarm Options
The alarm options associated with a parameter dictate how the system responds when an 
alarm occurs.
Bedside Alarm Tracking
Bedside Alarm Tracking provides an automatic bi-directional flow of alarm limit change and 
alarm delay information, between a patient bedside monitor and a Panorama Central 
Station. 
Latching
Alarm latching denotes that the audio and visual indicators, associated with an alarm 
condition, will not automatically stop even when the patient alarm condition is resolved. 
By default, lethal alarm events and nurse call events are automatically latched. Physiological 
alarm events can be configured to be latched or unlatched, and non-lethal alarms cannot be 
latched. When permitted, this option can be enabled/disabled in the System Alarms 
(System Alarm Options) tab.
The patient tile MUTE button temporarily silences the alarm condition for the configured time 
period. However, if that latched alarm is not acknowledged during the configured time 
period, the alarm indicators resume. Once a patient alarm condition is resolved, it must be 
acknowledged at the central station and at the bedside monitor. If it is not acknowledged, all 
audio and visual alarm indicators continue at the central station and at the bedside monitor, 
even if the alarm condition is resolved.
To acknowledge a latched alarm condition at the central station, once the condition is 
resolved, perform one of the following to the affected patient tile:
Select the digital data area 
Select the waveform area 
Select the VIEW button 
NOTE:
Acknowledging a latched alarm at the host central station 
patient tile acknowledges the alarm at the associated 
ViewStation and WorkStation patient tile.
NOTE:
Acknowledging a latched alarm at the bedside monitor 
does not acknowledge the alarm at the host central station, 
ViewStation, or WorkStation patient tile.
To acknowledge a latched alarm at a Passport 2 and/or Spectrum bedside monitor, refer to 
the Passport 2 and/or Spectrum Operating Instructions.