CARDIOVASCULAR SYSTEMS INC. EX1000 Manual De Usuario

Descargar
Página de 60
 
41 
 
Issue 
number 
Issue 
Solution 
on the OAS pump.  Verify that the OAS pump powers 
on. 
The OAS pump was 
running, but has 
stopped pumping 
and the yellow LED 
is illuminated 
1.  Verify the OAD power cord is connected to the OAS 
pump.  
2.  Press the Master Power switch, on the back of the 
OAS pump, to power off the OAS pump.  Wait five (5) 
seconds and press the Master Power switch to 
power on the pump. 
All three LEDs on 
the front panel of 
the OAS pump 
remain illuminated 
1.  Press the Master Power switch, on the back of the 
OAS pump, to power off the OAS pump.  Wait a few 
seconds and press the Master Power switch to 
power on the pump. 
2.  Contact CSI Customer Service at the phone number 
on the back of this instructions for use. 
After OAS pump 
power up, all three 
LEDs on the front 
panel of the OAS 
pump blink three 
times and there is 
an audible 
notification signal 
three times. 
1.  Contact CSI Customer Service at the phone number 
on the back of this instructions for use. 
The low saline level 
sensor (red LED) is 
illuminated 
Note: The OAS pump will stop pumping saline and 
supplying power to the OAD 30 seconds after the low 
saline level sensor activates while the OAD is 
spinning, as indicated by an audible information 
signal every 5 seconds. 
1.  If there is less than 200 mL of saline left in the bag of 
saline and lubricant, replace the bag with a new 1000 
mL bag of normal saline and lubricant solution. 
2.  Ensure that the bag of saline and lubricant is hanging 
freely from the saline bag open hook and that the low 
saline level sensor cord is properly inserted into the 
connector on the sensor and the connector on the 
back of the OAS pump. 
3.  Verify that the red low saline LED on the OAS pump 
control panel turns off and either the yellow LED or 
the green LED illuminates. If the yellow LED