CARDIOVASCULAR SYSTEMS INC. EX1000 Manual De Usuario
41
Issue
number
Issue
Solution
on the OAS pump. Verify that the OAS pump powers
on.
on.
3
The OAS pump was
running, but has
stopped pumping
and the yellow LED
is illuminated
running, but has
stopped pumping
and the yellow LED
is illuminated
1. Verify the OAD power cord is connected to the OAS
pump.
2. Press the Master Power switch, on the back of the
OAS pump, to power off the OAS pump. Wait five (5)
seconds and press the Master Power switch to
power on the pump.
seconds and press the Master Power switch to
power on the pump.
4
All three LEDs on
the front panel of
the OAS pump
remain illuminated
the front panel of
the OAS pump
remain illuminated
1. Press the Master Power switch, on the back of the
OAS pump, to power off the OAS pump. Wait a few
seconds and press the Master Power switch to
power on the pump.
seconds and press the Master Power switch to
power on the pump.
2. Contact CSI Customer Service at the phone number
on the back of this instructions for use.
5
After OAS pump
power up, all three
LEDs on the front
panel of the OAS
pump blink three
times and there is
an audible
notification signal
three times.
power up, all three
LEDs on the front
panel of the OAS
pump blink three
times and there is
an audible
notification signal
three times.
1. Contact CSI Customer Service at the phone number
on the back of this instructions for use.
6
The low saline level
sensor (red LED) is
illuminated
sensor (red LED) is
illuminated
Note: The OAS pump will stop pumping saline and
supplying power to the OAD 30 seconds after the low
saline level sensor activates while the OAD is
spinning, as indicated by an audible information
signal every 5 seconds.
saline level sensor activates while the OAD is
spinning, as indicated by an audible information
signal every 5 seconds.
1. If there is less than 200 mL of saline left in the bag of
saline and lubricant, replace the bag with a new 1000
mL bag of normal saline and lubricant solution.
mL bag of normal saline and lubricant solution.
2. Ensure that the bag of saline and lubricant is hanging
freely from the saline bag open hook and that the low
saline level sensor cord is properly inserted into the
connector on the sensor and the connector on the
back of the OAS pump.
saline level sensor cord is properly inserted into the
connector on the sensor and the connector on the
back of the OAS pump.
3. Verify that the red low saline LED on the OAS pump
control panel turns off and either the yellow LED or
the green LED illuminates. If the yellow LED
the green LED illuminates. If the yellow LED