CARDIOVASCULAR SYSTEMS INC. EX1000 Manual De Usuario
43
Issue
number
Issue
Solution
12
Crown rotational
speeds are variable
and will not stabilize
speeds are variable
and will not stabilize
1. Immediately discontinue treatment, but leave the
OAS pump running.
2. Verify that saline is flowing. Verify VIPERSLIDE
Lubricant is present in the saline bag. See the
VIPERSLIDE Lubricant Instructions for Use for
information.
VIPERSLIDE Lubricant Instructions for Use for
information.
3. Verify that the saline tubing is properly connected to
the saline bag, that the saline tubing is routed
correctly through the OAS pump saline tubing guides,
and that the saline tubing is properly connected to
the OAD.
correctly through the OAS pump saline tubing guides,
and that the saline tubing is properly connected to
the OAD.
4. Verify that the crown advancer knob moves
smoothly.
5. Retract the crown proximal to the lesion. Using a
travel rate between 1 mm per second and 10 mm per
second, continue treatment on low speed.
second, continue treatment on low speed.
6. If rotational speeds will not stabilize, replace the OAD
or guide wire.
13
The crown is not
moving one-to-one
with the crown
advancer knob
moving one-to-one
with the crown
advancer knob
During start up in the vessel:
1. Verify the Tuohy valve is not over-tightened.
2. Verify the crown advancer knob moves smoothly.
3. Retract the crown advancer knob until the crown
1. Verify the Tuohy valve is not over-tightened.
2. Verify the crown advancer knob moves smoothly.
3. Retract the crown advancer knob until the crown
moves with the knob.
While spinning:
1. Immediately discontinue treatment, but leave the
OAS pump running.
2. Verify the Tuohy valve is not over-tightened.
3. Verify that the crown advancer knob moves
3. Verify that the crown advancer knob moves
smoothly.
4. Retract the crown advancer knob until the crown
moves with the knob.
5. Verify that contrast media injections were not above
400 psi or occurred during crown spinning.
6. Engage and disengage the lesion using a travel rate
between 1 mm per second and 10 mm per second
while maintaining one-to-one crown to advancer knob
movement.
while maintaining one-to-one crown to advancer knob
movement.
14
The crown stops
spinning during the
procedure
spinning during the
procedure
1. Immediately discontinue treatment, but leave the
OAS pump running.
2. Check to ensure that the OAS pump power cord is
connected to the back of the OAS pump and that the
OAD power cord is connected to the OAS pump.
OAD power cord is connected to the OAS pump.