CARDIOVASCULAR SYSTEMS INC. EX1000 Manual De Usuario

Descargar
Página de 60
 
43 
 
Issue 
number 
Issue 
Solution 
12 
Crown rotational 
speeds are variable 
and will not stabilize 
1.  Immediately discontinue treatment, but leave the 
OAS pump running. 
2.  Verify that saline is flowing. Verify VIPERSLIDE 
Lubricant is present in the saline bag. See the 
VIPERSLIDE Lubricant Instructions for Use for 
information.  
3.  Verify that the saline tubing is properly connected to 
the saline bag, that the saline tubing is routed 
correctly through the OAS pump saline tubing guides, 
and that the saline tubing is properly connected to 
the OAD.  
4.  Verify that the crown advancer knob moves 
smoothly. 
5.  Retract the crown proximal to the lesion.  Using a 
travel rate between 1 mm per second and 10 mm per 
second, continue treatment on low speed. 
6.  If rotational speeds will not stabilize, replace the OAD 
or guide wire. 
13 
The crown is not 
moving one-to-one 
with the crown 
advancer knob 
During start up in the vessel: 
1.  Verify the Tuohy valve is not over-tightened. 
2.  Verify the crown advancer knob moves smoothly. 
3.  Retract the crown advancer knob until the crown 
moves with the knob. 
 
While spinning: 
1.  Immediately discontinue treatment, but leave the 
OAS pump running. 
2.  Verify the Tuohy valve is not over-tightened. 
3.  Verify that the crown advancer knob moves 
smoothly. 
4.  Retract the crown advancer knob until the crown 
moves with the knob. 
5.  Verify that contrast media injections were not above 
400 psi or occurred during crown spinning. 
6.  Engage and disengage the lesion using a travel rate 
between 1 mm per second and 10 mm per second 
while maintaining one-to-one crown to advancer knob 
movement. 
14 
The crown stops 
spinning during the 
procedure 
1.  Immediately discontinue treatment, but leave the 
OAS pump running. 
2.  Check to ensure that the OAS pump power cord is 
connected to the back of the OAS pump and that the 
OAD power cord is connected to the OAS pump.