Troubles
Possible causes
Solutions
6. The MStim Arth APP
on the mobile phone pop
up tips
on the mobile phone pop
up tips
“electrodes fall
off
”, the indictor of the
unit flashed Yellow.
1. The electrode lead hose
connected badly;
2. The electrodes are poor
contact with the skin.
3. Electrodes do not stick well;
connected badly;
2. The electrodes are poor
contact with the skin.
3. Electrodes do not stick well;
1. Re-connect the electrode lead
hose;
2. Re-connect the electrodes. 3.
Replace
hose;
2. Re-connect the electrodes. 3.
Replace
electrodes.
Apply
electrodes to a clean, dry
surface.
surface.
7. The unit has output,
but treatment without
sensation.
but treatment without
sensation.
The output intensity is too low.
Increase the output intensity.