GE Inspection Technologies LP 7260H Manual De Usuario

Descargar
Página de 83
25
07/30/2014
1
 – Tap the lower-left corner of the display (typically contains a GE Logo) or (or press 
the 
 hard key) at any time to open the Global Menu, which provides access to 
the Setup Menu.  
2
 – Select to alter the Measurement and Annotation-specific settings shown here.  
3
 – View the list of 3DPM Tips already calibrated for use with the attached probe. 
4
 – Select to view a list of Stereo Tips already-calibrated for use with the attached 
probe. The Add Tip and Edit List options are not intended for general use. Deleting 
a tip through the Edit List option will permanently remove that tip’s calibration data 
requiring the probe and tip to be returned to an authorized GE Inspection Technologies 
service center for re-calibration.
5
 – Save Preset Annotation (notes) to an external storage device or Import preset notes 
to the instrument from an external device.
6
 – Displays or removes the Zoom Window (used for detailed cursor placement during 
measurement). 
7
 – Select AUTO to show the 3DPM red/yellow un-measurable pixel mask only when a 
cursor is placed on a red/yellow pixel. Select ON to always show the pixel mask when 
measuring.  
8
 – Select Z-Map for most usage. P-Map images contain additional data and may be 
requested by GE Technical Support to aid in troubleshooting.
9
 – Controls the display of the Stereo Index, which appears during Stereo Measurement. 
10
 – Sets a threshold value for the Stereo Index below which the index flashes during 
Stereo Measurement.  This may be used to warn the user of low-index measurements 
where the accuracy may be reduced.
11
 – Specify the unit of measurement as Inches or Millimeters.