BMC Medical Co. Ltd. YH600AYH600B Manual De Usuario
YH‐600B Pro User Manual V1.0
17 / 49
When recording starts, the device can
detect the expansion module kit, but
fails to collect airflow waveform
information
fails to collect airflow waveform
information
Wear the Airflow Cannula
properly or long-press the
On-Off/OK Button
to
start recording compulsorily
When recording starts, the device can
detect the airflow kit, but fails to collect
thoracoabdominal respiration
information
thoracoabdominal respiration
information
Wear the respiration effort
belt properly or long-press the
On-Off/OK Button
to
start recording compulsorily
The battery is low when recording
starts
Replace the battery or press
and hold the On-Off/OK
Button
for
three
seconds to enter the “Record”
interface
interface
10.2.3.2 “Record” Interface
If there are no wearing alerts, the “Record” interface can be accessed directly. The time notification
(0:00:00) in the upper left corner of the “Record” interface flickers. Press and hold the On-Off/OK Button
(0:00:00) in the upper left corner of the “Record” interface flickers. Press and hold the On-Off/OK Button
for two seconds to start timing and recording monitored data in the device.
If there is no operation within the preset screen backlight timeout (except when the screen is set to stay
bright all the time), the screen will turn off automatically. Pressing any key can light up the screen again.
bright all the time), the screen will turn off automatically. Pressing any key can light up the screen again.
When recording is completed, press the On-Off/OK Button
and Shift Button
simultaneously
to exit the “Record” interface and return to the standby interface.
The “Record” interface of YH-600B Pro, as shown in Figure 1-12.
Figure 1-12 “Record” interface of YH-600B Pro
SpO
2
(%):
Blood oxygen saturation expressed in percentage
PR(bpm):
Pulse rate expressed as beats per minute (bpm)
Recording time:
The length of a recording
Snore:
Waveform indicates the occurrence of a snore