BMC Medical Co. Ltd. YH600AYH600B Manual De Usuario

Descargar
Página de 53
YH‐600B Pro User Manual V1.0 
  29 / 49 
determined by an electromagnetic site survey,   
a
 should be less than the compliance level in each 
frequency range.   
b
 Interference may occur in the vicinity of equipment 
marked with the following symbol:   
   
Note 1:    At 80~800 MHz, the higher frequency range applies. 
Note 2:  These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people. 
a
 
 
Field strength from fixed transmitters such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the device is used exceeds 
the applicable RF compliance level above, the device should be observed to verify normal 
operation. If abnormal performance is observed, additional measures may be necessary, such 
as re-orienting or relocating the device. 
b
 
Over the frequency range 150 kHz~  80 MHz, field strengths should be less than 3 V/m. 
 
Recommended Separation Distance Between Portable and Mobile RF Communication Equipment and the 
Device. 
The device is intended for use in the electromagnetic environment in which radiated RF disturbances are 
controlled. The user of the device can help prevent electromagnetic interference by maintaining a minimum 
distance between portable and mobile RF communication equipment (transmitters) and the device as 
recommended below, according to the maximum output power of the communication equipment. 
Rated Maximum Output Power of 
Transmitter /W 
Separation Distance According to Frequency of Transmitter/m   
150 kHz~80 MHz 
 
80~800 MHz 
 
800 MHz~2.5 GHz 
 
0.01 0.12 
0.12 
0.23 
0.1 0.38 
0.38 
0.73 
1 1.2 
1.2 
2.3 
10 3.8 
3.8 
7.3 
100 12 
12 
23 
For transmitters rated at a maximum output power not listed above, the recommended separation distance d 
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is 
the maximum output power of the transmitter manufacturer. 
Note 1:    At 80~800 MHz, the higher frequency range applies. 
Note 2:  These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.