YH‐600B Pro User Manual V1.0 
  27 / 49 
14. EMC Requirements   
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions 
This device is intended for use in the electromagnetic environment specified below. The user of the device 
should assure that it is used in such an environment. 
Emissions Test 
Compliance 
Electromagnetic Environment - Guidance 
RF Radiated Emissions 
CISPR 11 
Group 1 
The device uses RF energy only for its internal function. 
Therefore, its RF radiated emissions are very low and are 
not likely to cause any interference in nearby electronic 
equipment 
RF Conducted Emissions 
CISPR 11 
Class B 
The device is suitable for use in all establishments 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes 
Harmonic Emissions 
IEC 61000-3-2 
N/A 
Voltage Fluctuations/Flicker 
Emissions IEC 61000-3-3 
N/A 
 
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity 
This device is intended for use in the electromagnetic environment specified below. The user of the device 
should assure that it is used in such an environment. 
Immunity Test 
IEC 60601 Test 
Level 
Compliance Level
Electromagnetic Environment - Guidance 
Electrostatic 
Discharge 
(ESD) 
IEC 61000-4-2 
±6 kV Contact 
±8 kV Air 
±6 kV Contact 
±8 kV Air 
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30% 
Electrical Fast 
Transient/Burst 
IEC 61000-4-4 
±2 kV for   
power supply lines 
±1 kV for   
input/output lines 
N/A 
Mains power quality should be that of a 
typical commercial or hospital environment 
Surge  
IEC 61000-4-5 
±1 kV   
differential mode 
±2 kV   
common mode 
N/A 
Mains power quality should be that of a 
typical commercial or hospital environment 
Voltage Dips, 
Short 
Interruptions 
and Voltage 
Variations on 
Power Supply 
<5 % U
T
   
(>95% dip in U
T
)  
for 0.5 cycle   
40% U
T
   
(60% dip in U
T
)  
N/A 
Mains power quality should be that of a 
typical commercial or hospital environment. 
If the user of the device requires continued 
operation during power mains interruptions, 
it is recommended that the device be 
powered from an uninterruptible power