Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manual De Usuario

Los códigos de productos
MSO1074Z-S
Descargar
Página de 172
RIGOL 
DS1000Z Programming Guide 
2-87
 
:TRIGger:PULSe 
 
Command List: 
 
:TRIGger:PULSe:SOURce 
Syntax 
:TRIGger:PULSe:SOURce <source> 
:TRIGger:PULSe:SOURce? 
Description 
Set or query the trigger source in pulse width trigger.   
Parameter 
Name 
Type 
Range 
Default 
<source>  Discrete 
{CHANnel1|CHANnel2|CHANnel3|CHANnel4} 
CHANnel1 
 
Return 
Format 
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.   
Example 
:TRIGger:PULSe:SOURce CHANnel1    /*Set the trigger source to CH1*/ 
 
:TRIGger:PULSe:WHEN 
Syntax 
:TRIGger:PULSe:WHEN <when> 
:TRIGger:PULSe:WHEN? 
Description 
Set or query the trigger condition in pulse width trigger. 
 
Parameter 
Name 
Type 
Range 
Default 
<when>  Discrete  {PGReater|PLESs|NGReater|NLESs|PGLess|NGLess}  PGReater 
 
Explanation 
  PGReater/PLESs: you need to specify a pulse width (refer 
to 
). The oscilloscope triggers when the positive pulse width 
of the input signal is greater/lower than the specified Pulse Width. 
  NGReater/NLESs: you need to specify a pulse width (refer 
to 
). The oscilloscope triggers when the negative pulse width 
of the input signal is greater/lower than the specified Pulse Width. 
  PGLess/NGLess: you need to specify a upper (refer to 
and 
a lower (refer to 
)pulse width. The oscilloscope triggers 
when the positive/negative pulse width of the input signal is greater than the 
specified lower pulse width and lower than the upper pulse width.
 
Return 
Format 
The query returns PGR, PLES, NGR, NLES, PGL or NGL.   
Example 
:TRIGger:PULSe:WHEN NLESs    /*Set the trigger condition to NLESs*/ 
Related 
Command