Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manuel D’Utilisation
Codes de produits
MSO1074Z-S
RIGOL
DS1000Z Programming Guide
2-87
:TRIGger:PULSe
Command List:
:TRIGger:PULSe:SOURce
Syntax
:TRIGger:PULSe:SOURce <source>
:TRIGger:PULSe:SOURce?
:TRIGger:PULSe:SOURce?
Description
Set or query the trigger source in pulse width trigger.
Parameter
Name
Type
Range
Default
<source> Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:TRIGger:PULSe:SOURce CHANnel1 /*Set the trigger source to CH1*/
:TRIGger:PULSe:WHEN
Syntax
:TRIGger:PULSe:WHEN <when>
:TRIGger:PULSe:WHEN?
:TRIGger:PULSe:WHEN?
Description
Set or query the trigger condition in pulse width trigger.
Parameter
Name
Type
Range
Default
<when> Discrete {PGReater|PLESs|NGReater|NLESs|PGLess|NGLess} PGReater
Explanation
PGReater/PLESs: you need to specify a pulse width (refer
). The oscilloscope triggers when the positive pulse width
of the input signal is greater/lower than the specified Pulse Width.
NGReater/NLESs: you need to specify a pulse width (refer
). The oscilloscope triggers when the negative pulse width
of the input signal is greater/lower than the specified Pulse Width.
) and
)pulse width. The oscilloscope triggers
when the positive/negative pulse width of the input signal is greater than the
specified lower pulse width and lower than the upper pulse width.
Return
Format
The query returns PGR, PLES, NGR, NLES, PGL or NGL.
Example
:TRIGger:PULSe:WHEN NLESs /*Set the trigger condition to NLESs*/
Related
Command