Benning IT 120 B 044102 Manual De Usuario

Los códigos de productos
044102
Descargar
Página de 80
51
Fig. 6.5: Recall results menu
 
 
 Use the 
 keys to select the function for which you want to view results. Press the 
TEST key to confirm.
 
Fig. 6.6: Examples of recall results menu
 
 
By using the 
 keys other saved results can be viewed under the same object, 
block and fuse items. Press 
 or 
 to return to RECALL RESULTS menu.
Note: 
No measurement results can be recalled if the switch is set to „V~“.
Exit the memory recall: Use the „MEM“ key (several times) or the selector switch.
6.3  Deleting measurement results
When deleting results the following actions can be taken:
-  Individual results can be deleted
-   Results under the same structure item can be deleted
-   All saved results can be deleted
-   Exit the menu by turning the selector switch or by using the „MEM“ key
How to delete individual saved results:
Step 1
 
 To enter the Memory menu press the MEM key.
 
 
 By using the 
 keys set cursor to DELETE RESULTS. Press the TEST key, the fol-
lowing menu is displayed: 
Fig. 6.7:  Delete results menu
Step 2 
By using the 
 keys set cursor to OBJECT line.
 
 
Use the 
 keys to enter the desired element OBJECT xx.
 
 
By using the 
 keys set cursor to BLOCK line. 
 
 
Use the 
 keys to enter the desired element BLOCK xx.
 
 
By using the 
 keys set cursor to FUSE line. 
 
 
Use the 
 keys to enter the desired element FUSE xx.
 
 
In No. line the number of saved results is displayed.