CH 4/ Instrument Settings
Touch Screen Calibration
Calibration 
procedure
The Energy Platform is able to perform calibration to ensure the proper operation of the 
unit’s touch screen functionalities. The calibration procedure will correct the problem 
of out of tolerance touch screen malfunction. Note that errors corrected by this 
calibration procedure are specific only to touch screen operation. 
Continued on next page
Action...
Result...
STEP 1: From the Display Setup 
Menu (see page 4-6) screen, press 
Calibrate Touch Screen.Test if 
touch screen recognition operates 
properly by applying pressure on 
the LCD screen. 
• Press Clear to reset LCD 
screen and delete display 
markings.
• Press Calib to begin touch 
screen calibration procedure. 
Proceed to Step 2.
• Press Cancel to end calibration 
test and return to Display 
Setup.
MARK164b
STEP 2: Follow the instruction 
Tap Center of Target 
to begin 
calibration. Target object is 
initially located in the lower 
middle section of the screen. 
• A series of screens will flash 
showing movement of the 
target object: from the lower 
middle section to mid-right and 
finally to upper left section of 
the screen.
• To end touch screen 
calibration, tap the center of the 
target object X now located in 
the upper left section of the 
screen. Proceed to Step 3 on 
page 4-8.
MARK164
MARK165
MARK166