Touch Screen Calibration, continued
Calibration 
procedure 
(continued)
Action...
Result...
STEP 3: Once touch screen 
calibration is done, a message will 
appear to indicate Touch screen 
was successfully calibrated. 
• Press OK to continue on with 
calibration test verification. 
Proceed to Step 4.
MARK167
STEP 4: Test if touch screen 
recognition operates properly by 
applying pressure on the LCD 
screen. 
• Press Clear to reset LCD 
screen and delete display 
markings.
• Press Accept to complete the 
touch screen calibration 
procedure. Accepting will save 
and store new touch screen 
calibration data in memory. 
Once the new data is stored, the 
old touch screen calibration 
data is lost. The screen will 
return to the Display Setup 
Menu.
• Press Cancel to discontinue the 
touch screen calibration 
procedure. Pressing Cancel will 
retain the previous touch screen 
calibration constants, no new 
calibration data will be stored 
in memory. The screen will 
return to the Display Setup 
menu.
MARK168