CH 5/ Start Menu
Scale factor 
setup screen
After setting up current probe configurations, the Scale Factor screen is displayed.
Action...
Result...
STEP 1: From the Current Probe 
Selection screen, press Next to 
display the Scale Factor Setup.
• If voltage and current 
connections are made without 
using any current or potential 
transformer devices, the scale 
factor is set to the default value 
of 1. Press Next and go to page 
5-12.
• To enter new scale factors, 
press the numeric field of the 
Volts/Amps channel where you 
want to enter new scale factors. 
Proceed to Step 2.
• Press Cancel to quit and return 
to Start menu.
MARK105
STEP 2: Enter a new scale factor 
value using the numeric keypad. 
Press OK to accept the new scale 
factor.
• Press 
.. .
 
. . 
to clear and retype 
numeric values.
• Press x1000 to multiply the 
value by 1000.
• Press Cancel to discard 
changes and retain previous 
scale factor value, and return to 
Scale Factor Setup screen.
MARK105b