5-12
Circuit Type Selection
Circuit Type Selection
Circuit setup
 
The Energy Platform is able to display wiring diagrams from which users can select the 
circuit type appropriate to their application. The instrument will display the detected 
circuit configuration and compare the detected voltages, currents, and phase sequence 
(if applicable) to the selected circuit type. Refer to Appendix E for the descriptions of 
common circuit connections. 
Continued on next page
Action...
Result...
STEP 1: From the Scale Factor 
Setup screen, press Next to 
display Circuit Type Selection. 
Read the Warning advisory, then 
press Accept to display the menus 
for circuit selection.
• To select a circuit 
configuration, press the Ckt 
arrow keys until the desired 
configuration is displayed.
An auto circuit detection window 
is displayed to show if selected 
circuit type and the detected 
voltages, currents, and phase 
sequence match (text in green) or 
do not match (text in red). 
When done with Circuit setup, 
press Next. Go to page 5-14.
• Real-time Scope and Phasor 
screens are available. These 
functions help verify if the 
wiring configuration matches 
the circuit being monitored.
Press Scope to check if the 
waveforms look correct for the 
specified wiring setup. Proceed 
to Step 2 on page 5-13.
Press Phasor to view if phasor 
display match those expected 
from the specified wiring 
diagram. Proceed to Step 3 on 
page 5-13.
• Press Cancel to discard 
changes and return to Start 
menu.
 
EP107
EP108