CH 4/ Instrument Settings
Touch Screen Calibration
Calibration
procedure
procedure
The Energy Platform is able to perform calibration to ensure the proper operation of the
unit’s touch screen functionalities. The calibration procedure will correct the problem
of out of tolerance touch screen malfunction. Note that errors corrected by this
calibration procedure are specific only to touch screen operation.
Continued on next page
Action...
Result...
STEP 1: From the Display Setup
Menu (see page 4-6) screen, press
Calibrate Touch Screen.Test if
touch screen recognition operates
properly by applying pressure on
the LCD screen.
• Press Clear to reset LCD
screen and delete display
markings.
• Press Calib to begin touch
screen calibration procedure.
Proceed to Step 2.
• Press Cancel to end calibration
test and return to Display
Setup.
MARK164b
STEP 2: Follow the instruction
Tap Center of Target
Tap Center of Target
to begin
calibration. Target object X is
initially located in the lower
middle section of the screen.
• A series of screens will flash
showing movement of the
target object: from the lower
middle section to mid-right and
finally to upper left section of
the screen.
• To end touch screen
calibration, tap the center of the
target object X now located in
the upper left section of the
screen. Proceed to Step 3 on
page 4-8.
MARK164
MARK165
MARK166