Voltcraft SL-451 Sound Level Meter and Data Logger 31.5 Hz – 8 kHz SL-451 Hoja De Datos

Los códigos de productos
SL-451
Descargar
Página de 2
 interface protocol
 
Interface protocol modification  
             9600 bit, with start and stop bit, 8 bit transmitting, without parity check.  
The communication protocol from SCM to PC 
Protocol 1.: data transmitting protocol 
Identification 
code 
 data 
hex BCD 
code 
A5 0D  
 
A5 06     
 
The first byte fixed at A5, which is data identification code. 
The following is BCD code for 2 byte(data display) or 3 byte(time). 
Time style 0A5H + 06H + 3 byte data   
2 kinds of data style Renovating data style ---0A5H + 0DH + 2 byte data + A5H 
  + 0BH 
 
 
 
 
Renovating bargraph style ---0A5H + 0DH    + 
2 byte data +  A5H   +  0CH
 
The indicated bargraph segments are calculated according to 2 byte data and current 
range.( the last bit of BCD is decimal fraction). 
  Command and function  
 hex 
indicator Data 
range 
Available 
reading 
 
A5H 
1BH 
dBA 
0-130.0 dB 
0.1 dB 
A5H 
1CH 
dBC 
0-130.0 dB 
0.1 dB 
dBA / dBC only one can be 
displayed 
A5H 02H 
FAST 
A5H 03H 
SLOW 
FAST/SLOW only one can be displayed 
A5H 04H 
MAX 
A5H 05H 
MIN 
MAX/MIN only one can be displayed or non of them 
displayed 
A5H  06H 
TIME 
1:00:00—12:59:59   Display and renovating time, date, 
year, month etc. 
A5H 
07H 
OVER 
Display OVER bargraph and current measured readings 
A5H 08H 
UNDER 
 
A5H 
0ffH 
hold 
PC only receive command but not display HOLD symbol 
A5H 
0AH 
REC 
Automatically Saving function  A5H  1AH  cancel REC  
A5H 
0BH 
Display renovating data and bargraph 
A5H 
0CH 
Display renovating bargraph but not data 
A5H 
0DH 
Display measured readings with decimal all the time. 
A5H 0EH 
CANCEL 
MAX/MIN 
A5 
11H 
CANCEL OVER& UNDER