Voltcraft SL-451 Sound Level Meter and Data Logger 31.5 Hz – 8 kHz SL-451 Hoja De Datos

Los códigos de productos
SL-451
Descargar
Página de 2
  
range 
A5H 10H 
30 
Db-----80 
dB 
A5H 
20H 
50 dB-----100 dB 
A5H 
30H 
80 dB-----130 dB 
A5H 
40H 
30 dB-----130dB    display AUTO at the same time 
 
Protocol 2 :   
In PC interface protocol, the received data should be managed as following: 
1. according to received data and measured range, calculating and display the bargraph segements 
(bargraph segements are 51)      
2.  Switchable received data at anytime, when to receive the data, displaying the Max/Min noise, 
average value and opposite time, with automatic saving functiuon. 
3.  Input the data to access or excel table 
4.  change it to graph 
5. 
choosing COM jack, bit, start bit and 8 bit transmitting.
 
6.  Transmitting 5AH    0ACH reading  DATA LOGGER receiving    0DDH 
start 
data 
analyzing 
with saving or not function  
The communication protocol from SCM to PC 
The command style of PC: 
Command and function: 
 
command function 
33H Power 
off 
55H 
Send out rec command cancel rec command  
11H display 
max display min cancel MAX/MIN  
77H Display 
FAST display SLOW  
88H Range 
Switch 
 
99H 
Send out dBA command send out dBC command  
0ACH 
DATA  LOGGER read function 
DATA  LOGGER protocol 
1. data receiving 
BBH 
XH XL aa/cc year
month
date
hour
minute
Sec.
Sampling 
rate 
ACH
data
BBH Start 
signal 
 
XH/XL Data 
volume 
AAH/CCH DBA&DBC 
ACH 
Start receiving data 
DDH Over 
symbol