Gw Instek AFG-2025 Function Generator, Frequency Generator AFG-2025 Hoja De Datos

Los códigos de productos
AFG-2025
Descargar
Página de 145
 
 
AFG-2000 Series User Manual 
96
 
Apply Commands 
 
The APPLy command has 5 different types of outputs (Sine, Square, 
Ramp, Noise, User(ARB)). The Apply command is the quickest, 
easiest way to output waveforms remotely. Frequency, amplitude 
and offset can be specified for each function.  
As only basic parameters can be set with the Apply command, other 
parameters, such as duty and symmetry use the instrument default 
values. 
The Apply command will set the trigger source to immediate and 
disable modulation and sweep modes, if active. The command also 
turns on the output command SOURce[1]:OUTP ON. 
 
As the frequency, amplitude and offset parameters are in nested 
square brackets, the amplitude can only be specified if the frequency 
has been specified and the offset can only be specified if the 
amplitude has been set.  See the syntax below for the example: 
 
SOURce1:APPLy:<function> [<frequency> [,<amplitude> 
[,<offset>] ]] 
Output Frequency 
For the output frequency, MINimum, MAXimum 
and DEFault can be used instead of specifying a 
frequency. The default frequency for all functions 
is set to 1 kHz.  
 
The maximum and minimum frequency depends 
on the function used and the model of the 
frequency generator.  If a frequency output that is 
out of range is specified, the max/min frequency 
will be used instead. A ―-222‖ error will be 
generated from the remote terminal.  
 
Function 
Min frequency 
Max frequency 
Sine 
0.1Hz 
25MHz*