MD 9030 TRMS Electrical & Electronic DMM with Temperature
Operation
8
CAUTION
Using Resistance, Continuity, Diode or Capacitance function in a live circuit will produce
false results and may damage the instrument. In many cases the suspected component
must be disconnected from the circuit to obtain an accurate measurement reading.
Using Resistance, Continuity, Diode or Capacitance function in a live circuit will produce
false results and may damage the instrument. In many cases the suspected component
must be disconnected from the circuit to obtain an accurate measurement reading.
Diode test,
Capacitance functions
Defaults at
. Press SELECT button momentarily 2 times to select
Diode test
function. Normal forward voltage drop (forward biased) for a good silicon diode is
between 0.400V to 0.900V. A reading higher than that indicates a leaky diode
(defective). A zero reading indicates a shorted diode (defective). An OL indicates an
open diode (defective). Reverse the test leads connections (reverse biased) across the
diode. The digital display shows OL if the diode is good. Any other readings indicate the
diode is resistive or shorted (defective).
between 0.400V to 0.900V. A reading higher than that indicates a leaky diode
(defective). A zero reading indicates a shorted diode (defective). An OL indicates an
open diode (defective). Reverse the test leads connections (reverse biased) across the
diode. The digital display shows OL if the diode is good. Any other readings indicate the
diode is resistive or shorted (defective).
Defaults at
. Press SELECT button momentarily 3 times to select
Capacitance
function. Relative zero
mode can be used to zero out the parasitic capacitance of the