MD 9010 General Purpose Auto – Check Digitalmultimeter Specification 
 
11 
DC Voltage (MD 9010
RANGE Accuracy 
6.000V 0.5%+3d 
60.00V 1.0%+5d 
450.0V
 1)
 1.2%+5d 
600.0V
 2)
 2.0%+5d 
600.0V
 3)
 1.5%+5d 
Input Impedance: 
AutoCheck
TM
 Lo-Z DCV: 833k
 (4.2k 
when displaying “Auto”), 90pF nominal 
Hi-Z DCV: 5M
, 90pF nominal 
NMRR: > 30dB @ 50Hz/60Hz 
CMRR: > 100dB @ DC, 50Hz/60Hz; 
Rs=1k
  
 
DCV AutoCheck
TM
 Threshold: 
> +1.2VDC or < -0.6VDC nominal 
2)
MD 9010 AutoCheck
TM
 DCV only 
3)
MD 9010 Hi-Z DCV only 
 
 
AC Voltage (MD 9010) 
RANGE Accuracy 
50Hz -- 60Hz 
6.000V, 60.00V, 
450.0V
 1)
, 600.0V
 2)
 
1.5%+5d 
CMRR: > 60dB @ DC to 60Hz, 
Rs=1k
 
MD 9010 Input Impedance: 
AutoCheckTM Lo-Z ACV: 833k
 
(4.2k
 when displaying “Auto”), 
90pF nominal 
Hi-Z ACV: 5M
, 90pF nominal 
ACV AutoCheck
TM
 Threshold: 
> 1.5VAC (50/60Hz) nominal 
2)
Range for MD 9010 only 
 
 
Capacitance (MD 9010) 
RANGE
 1)
 Accuracy
 2)
100.0nF, 1000nF, 
10.00
F, 100.0F
 3)
2000
F
 4)
 
3.5%+6d
 5)
 
1)
Accuracy below 50nF is not 
specified 
2)
Accuracies with film capacitor or 
better 
4)
MD 9010. Updates > 1 minute 
on large values 
5)
Specified with battery voltage 
above 2.8V (half full battery). 
Accuracy decreases gradually to 
12% at low battery warning 
voltage of approx 2.4V 
 
 
Ohms (MD 9010) 
RANGE
 1)
 Accuracy 
600.0
 
2.0%+6d
 2)
 
6.000k
 
1.2%+6d
 2
60.00K
, 600.0K 
1.0%+4d 
6.000M
 
2.0%+4d 
Open Circuit Voltage: 0.4VDC typical  
1)
AutoCheck
TM
 is for 6.000k
 ~ 6.000M 
ranges; 600.0
 is an independent range 
for MD 9010 only  
2)
Add 40d to specified accuracy while 
reading is below 20% of range 
 
 
Hz Frequency (MD 9010
 1) 2)
RANGE 
Accura
cy 
Specified at
10.00Hz -- 
30.00kHz
 1) 2)
 
< 20V Sine-
rms 
10.00Hz -- 
999.9Hz
 2)
 
0.5%+4
< 600V Sine-
rms 
Sensitivity (Sine-rms): 
1)
Hz in Auto-V
 position:  > 3V 
2)
Line-level Hz in V position (MD 9010 
only): 
@ 6.000VAC range: 
> 3V 
@ 60.00VAC range: 
> 6V 
@ 600.0VAC range: 
> 60V 
 
Diode Tester (MD 9010) 
Test Current 
(Typical) 
Open Circuit 
Voltage 
0.48mA for MD 
9010 
< 1.6VDC 
 
DC 
A Current (MD 9010) 
RANGE Accuracy 
Burden 
Voltage
400.0
A 
1.5%+3d 
6mV/
A 
2000
A 
1.2%+3d 
6mV/
A