Measurement Options 
43 
 
Table 4-1 Auto delay settings 
  Function 
Range and Delay 
DCV 
120mV 
1ms 
1.2V 
1ms 
12V 
1ms 
120V 
5ms 
1000V 
5ms 
 
 
 
 
ACV 
120mV 
400ms 
1.2V 
400ms 
12V 
400ms 
120V 
400ms 
750V 
400ms 
 
 
 
 
FREQ 
120mV 
1ms 
1.2V 
1ms 
12V 
1ms 
120V 
1ms 
750V 
1ms 
 
 
 
 
DCI 
12mA 
2ms 
120mA 
2ms 
1.2A 
2ms 
12A 
2ms 
 
 
 
 
 
 
ACI 
12mA 
400ms 
 
 
1.2A 
400ms 
12A 
400ms 
 
 
 
 
 
 
Ω2W, Ω4W 
12
0Ω 
3ms 
1.2
kΩ 
3ms 
12
kΩ 
13ms 
12
0kΩ 
25ms 
1.2
MΩ 
100ms 
12
MΩ 
150ms 
12
0MΩ 
250ms 
Continuity 
 
 
1kΩ 
3ms 
 
 
 
 
 
 
 
Diode testing 
 
1mA 
1ms 
100uA
 
1ms 
10uA 
1
ms 
 
 
 
 
To set the delay manually, follow the below steps: 
8.  Press 
  →         to enter the menu on the menu level, “A: MEAS MENU” will be displayed. 
9.  Use        or        key to move across to the TRIG MENU on the menu level, “C: TRIG MENU” will 
be displayed.   
10. Press         to move down to the command level within the TRIG MENU, “1: TRIG MODE” will be 
displayed. 
11. Use          to move across to the TRIG DELAY command on the command level, “2: TRIG DELAY” 
will be displayed. 
12. Press         to move down a level to set the type of delay mode. 
13. Using       or        to select AUTO or MANU (Manual) delay mode. 
14. Press 
(ENTER) to confirm the choice. The message “CHANGE SAVED” will be displayed to 
show that the change is now in effect.  If AUTO mode is selected,  the multimeter  will exit the 
parameter level and move up a level to the command level. 
15. If MANU mode is selected, a delay time will need to be specified.    The screen will display like below 
to show the current manual trigger delay value: 
DELAY:    0000mS 
16. Use        and        keys to choose a numerical place and use        and        keys to increment 
or decrement the digits. Enter a value for delay time (0 to 6000 ms). 
17. Press 
(ENTER) to confirm the delay time. The message “CHANGE SAVED” will be displayed 
to show that the change is now in effect. The multimeter will automatically exit the parameter level 
and move up a level to the command level. 
18. Press 
  →        key to exit from the menu and return to the measurement display. 
 
Note:    Changing the trigger delay to MANU (Manual) on any function changes the same for all 
other functions.