Measurement Options 
44 
 
4.2.2    EXT Trig & VM Comp 
The rear panel of the instrument has two BNC terminals:  Ext Trig and VM Comp.  This section 
describes the functionality of these two terminals. 
 
Ext Trig 
This is the external trigger input terminal.    When trigger mode is set to EXT, this terminal can be used 
as the source for triggering a measurement.  Input a negative pulse  into this terminal to trigger the 
instrument.    The pulse width 
must be greater than 1 μs.    An example is illustrated in Figure 4-3 below. 
 
5 V
0 V
Iutput
Ext  Trig
>
1
μs
 
Figure 4-3 
Trigger model 
 
VM Comp 
This is the voltmeter complete output terminal.  When the instrument completes  its  present 
measurement, a voltmeter complete output signal will output from this terminal as shown in Figure 4-4 
below. 
5 V
0 V
Output
VM Comp
Approximately
    2
μs
 
Figure 4-4 
Trigger model 
4.3    Buffer Operations 
The 5492B has an internal buffer to store 2 to 512 readings. In addition, the buffer includes statistical 
information  based on the stored readings  such as the  minimum  reading, maximum  reading,  average 
based on the stored readings, and standard deviation of the stored readings 
(See section “4.3.3     
 for details).
 
 
The buffer fills up with the requested number of readings and then stops.    Readings are then placed 
into the buffer and can be read back via front panel or remote operation.    Buffered data is overwritten 
each time the reading store function is turned enabled (ON) and initiated.    The data is volatile; it is not 
Input