Bk Precision BK-5492B Digital-Multimeter, DMM, BK-5492B Hoja De Datos

Los códigos de productos
BK-5492B
Descargar
Página de 111
Measurement Options 
49 
 
8.  Use        or        key to move across to the  LOW  LIMIT command, “8:  LOW  LIMIT”  will be 
displayed. 
9.  Press         to move down a level to input the low limit value. The current low limit value will be 
displayed: 
LO: -1.00000  
10.  Use       and        keys to choose a numerical place and use       and        keys to increment 
or decrement the digits. Enter a low limit value. 
11.  Press 
(ENTER)  to confirm  the value of low limit. The message “CHANGE SAVED”  will be 
displayed to show that the change is now in effect. 
12.  Press 
  →        key to exit from the menu. 
 
 
4.4.3    Configure Limit Beep 
Users can configure the multimeter to make a beep sound upon specified conditions when using limit 
operation.    Follow the below steps to configure: 
1.  Press 
  →        to enter the menu on the menu level, “A: MEAS MENU” will be displayed. 
2.  Use         or       key to move across to the MATH MENU on the menu level, “B: MATH MENU” 
will be displayed. 
3.  Press         to move down to the command  level within the MATH MENU, “1: SET M”  will be 
displayed. 
4.  Use        or       key to move across to the LIMIT BEEP  command,  “9:  LIMIT BEEP”  will be 
displayed. 
5.  Press          to move down a level to select the condition in which the meter should make a beep 
sound when using limit operation.    The display should show “ALARM:” with the current selection 
blinking. 
6.  Use        and        keys to choose between one of the following: 
NEVER:    Never beep regardless of the limit status (HI, LO, IN). 
HI: Beep only when the reading is HI (Reading is above the high limit). 
IN: Beep only when the reading is within the range specified by high and low limits. 
LO: Beep only when the reading is LO (Reading is below the low limit). 
OUT:    Beep only when the reading is HI or LO. 
7.  Press 
(ENTER) to confirm the value of high limit. The message “CHANGE SAVED” will be 
displayed to show that the change is now in effect. 
 
 
 
4.5    System Operations 
There are some system settings that can be configured on the 5492B multimeter, which include beeper 
control, saving and restoring instrument settings, front panel display control, key sound control, self-test, 
and calibration.    The following sections will describe in details each of these settings.