Measurement Options 
47 
 
4.3.3    Buffer Statistics 
In addition to measured readings that can be stored into the internal buffer, four other statistical 
information of the stored readings are also kept inside the buffer.    They are:    MAX, MIN, AVG, STD.     
 
MAX and MIN 
The MAX and MIN refer to the maximum and minimum value of the readings stored in the buffer 
respectively.     
 
For example, suppose the buffer contains the following five readings: 1.234, 2.345, 3.456, 4.567, 5.678. 
MAX = 5.678 
MIN = 1.234 
 
AVR 
AVR stands for average.    The AVR value is the mean(average) of the buffered readings. The equation 
used to calculate this is: 
�� =
��
��
��
��=1
��
 
Where: 
��
��
  is a stored reading 
       
��  is the number of stored readings 
 
For example, suppose the buffer contains the following five readings: 1.234, 2.345, 3.456, 4.567, 5.678. 
AVG = 3.456 
 
STD 
The STD value is the standard deviation of the buffered readings. The equation used to calculate the 
standard deviation is: 
�� = �
��
��
2
��
��=1
− �1���∑ ��
��
��
��=1
2
�� − 1
 
Where:   
��
��
  is a stored reading 
       
��  is the number of stored readings 
 
4.4    Limit Operations 
Limit operations set and control the values that determine the HI  /  IN  /  LO status of subsequent 
measurements. Limits can be applied to all measurement functions except continuity. The limit test  is 
performed after mX+b and percent math operations. Unit prefixes are applied before the limit test, for 
example: 
Low Limit = -1.0, High Limit = 1.0 
A 150mV reading becomes 0.15V (IN). 
 
Low Limit = -1.0, High Limit = 1.0