SCPI Command Reference 
94 
 
6.3.6    UNIT Subsystem 
The UNIT subsystem is used to configure and control the measurement units for ACV  and DCV, 
summarized in Table 6-6. 
Table 6-6 
UNIT Command Summary
 
Command 
Function description 
Default   
:UNIT 
  :VOLTage 
:AC <name> 
     :DB 
       :REFerence <n> 
       :REFerence? 
     :DBM 
       :IMPedance <n> 
       :IMPedance? 
    :AC? 
:DC <name> 
       :DB 
       :REFerence <n> 
       :REFerence? 
      :DBM 
       :IMPedance <n> 
       :IMPedance? 
    :DC? 
 
Path to configure voltage unit 
Select ACV measurement units (V, DB or DBM) 
Path to set DB reference voltage 
Specify reference in volts (1e-7 to 1000) 
Query DB reference 
Path to set DBM reference impedance 
Specify reference impedance (1 to 9999) 
Query DBM reference impedance 
Query ACV units 
Select DCV measurement units(V, DB, DBM) 
Path to set DB reference voltage: 
Specify reference in volts (0 to 1000) 
Query DB reference voltage 
Path to set DBM reference impedance: 
Specify reference impedance(1e-7 to 1000) 
Query DBM reference impedance 
Query DCV units 
 
 
 
 
 
75 
 
 
 
 
 
75 
 
 
 
:VOLTage:AC <name> 
Command Syntax: 
:UNIT:VOLTage:AC <name> 
 
Command parameter: 
<name> =    V    
 
 
AC voltage measurement units 
DB   
 
 
dB AC voltage measurement units 
DBM 
 
 
dBm AC voltage measurement units 
 
Query: 
:VOLTage:AC?             
Query AC voltage units 
Returned String: 
 
V, DB, or DBM 
 
Description: 
The command is used to select the units for ACV measurement. With volt (V) unit selected, normal 
AC  voltage measurement are made. With DB unit  selected, AC dB voltage measurements are 
performed.    With DBM units selected, AC dBm voltage measurements are made.