SCPI Command Reference 
95 
 
:DB:REFerence <n> 
Command Syntax: 
:UNIT:VOLTage:AC:DB:REFerence <n> 
 
Command Parameter: 
<n> =    1e-7 to 1000           
Specify reference voltage   
 
Query: 
      :DB:REFerence?             
Query reference voltage 
 
Description: 
This command is used to specify the dB reference level. When DB unit  is selected, ACV dB 
measurements are made using the specified reference voltage. 
The reference level is specified in volts and is not range dependent. For example: a dB reference 
level of 1 is 1 V on all ACV measurement ranges. 
 
:DBM:IMPedance <n> 
Command Syntax: 
:UNIT:VOLTage:AC:DBM:IMPedance <n> 
 
Command parameter: 
<n> = 1 to 9999                
 
Specify reference impedance 
 
Query: 
:DBM:IMPedance?             
Query reference impedance. 
 
Description: 
This command is used to specify the dBm reference impedance level. When dBm unit is selected, 
ACV dBm measurements are made using the specified reference impedance. 
 
The reference impedance is specified in ohms and is not range dependent. For Example: a dBm 
reference level of 500 is 500 on all ACV measurement ranges. 
 
:VOLTage:DC <name> 
Command Syntax: 
:UNIT:VOLTage:DC <name> 
 
Command Parameter: 
<name> =    V     
 
 
 
DC voltage measurement units 
 
 
 
 
DB   
      
 
dB DC voltage measurement units 
 
 
 
 
DBM 
 
 
 
dBm DC voltage measurement units 
 
Query: 
:VOLTage:DC?             
 
Query DC voltage units